VPASS Vaporized Pracetic Acid Decontamination System
VPASS is a decontamination system that vaporizes peracetic acid-based disinfectants and disperses them to every corner of a sealed space, leveraging the powerful sterilizing action of peracetic acid to decontaminate the interiors of laboratory equipment.
Because vaporized peracetic acid (VPA) can pass through HEPA filters, it can decontaminate the entire apparatus, including the HEPA filters themselves.
The system can be used for decontamination prior to the maintenance of biosafety cabinets (BSCs) or CO2 incubators, as well as for decontamination before removing BSCs from a facility.
Features
- Ensures safety using a low-concentration vaporized peracetic acid (approx. 30 ppm)
- Capable of achieving a 6-log or greater reduction of biological indicators
- Fully automated decontamination and chemical recovery processes minimize operational variance caused by operator experience
- The recirculation system design allows other tasks to be performed simultaneously in the same room during decontamination
- Eliminates the need for post-decontamination neutralization or prolonged aeration, thanks to chemical recovery via the filter unit
- Supports validation processes (CSV, IQ/OQ)
Overview of decontamination workflow
| Task details | Time taken | |
|---|---|---|
| Step1 | System delivery to the facility | Varies by facility |
| Step2 | System assembly | 15 minutes |
| Step3 | Attaching the exhaust adapter and masking glass surfaces | 30 minutes |
| Step4 | Bio-decontamination process | 2-3 hours Achieves a 6-log or greater reduction of biological indicators |
| Step5 | Bio-decontamination gas recovery | 30 minutes |
| Step6 | Removal of masking and exhaust adapter | 15 minutes |
| Step7 | System removal from the facility | Varies by facility |
| Total required working time | 3.5 - 4.5 hours | |
Peracetic acid vaporized inside the tank unit circulates throughout the target apparatus, decontaminating the entire interior, including the HEPA filter.
In the final process, the built-in filter unit removes the vaporized peracetic acid and reduces the acetic acid odor.
| Housing material | Staimless steel (SUS304) |
|---|---|
| Maximum external dimensions | 643(W)×662(D)×1300(H)mm Including protrusions, rear handles and warning lights. |
| *Actual simensions may vary depending on the actual unit. | |
| Operating mode | Fully automatic (Decontamination -> Gas recovery -> Chemical drainage) |
| Safety functions | Automatic blower shutdown (overheating, impeller lock), valve orientation erroe detection, door open detection, waste liquid bottle detection, emergency stop button |
| Gas leakage prevention | Sealing verification mechanism for decontamination targets and piping routes |
| Communication interface | Ethernet(Modbus TCP)、RS-485(Modbus RTU)、USB-A |
| Decontamination agent | Peracetic acid-based disinfectant (VPASS-dedicated agent, RTU) |
| Decontamination capacity | Proven: cabinet width up to 2150 mm |
| Operating temperature | 15 - 30 degC (validated environment) |
| Initial operating humidity | RH 45 - 75 % (validated environment) |
| Weight | Approx. 140 kg |
| Power supply | AC 100 - 240 V |
| Supported languages | English / Japanese (switchable) |